Efficient and reliable pure steam generators
Lautenschläger clean steam generators are specially designed for the high peak consumption of steam sterilizers and steam disinfectors. A large boiler volume, high working pressure and the low heating surface load ensure optimum dry saturated steam.
The Lautenschläger pure steam generator for optimum saturated steam
Technical data pure steam generator
Type | Heating | Heat output [kW] | Ø capacity [kg/h] | Tips up to [kg/1 min] | at max. [bar] |
---|---|---|---|---|---|
ED 15 | electric | 15 | 20 | 1 | 5 |
ED 36 | electric | 36 | 48 | 2,5 | 4 |
WT 54 | Heating steam | 54 | 70 | 4 | 5 |
ED 72 | electric | 72 | 96 | 5 | 5 |
WT 100 | Heating steam | 100 | 140 | 12 | 5 |
ED 144 | electric | 144 | 194 | 12 | 5/10 |
WT 100 S 200 | Heating steam | 200 | 280 | 12 | 4/10 |
WT 300 | Heating steam | 300 | 440 | 18 | 5 |
Pure steam for autoclaves in laboratories or in industrial and pharmaceutical applications must not be overheated or contain excessive moisture. It should also be as free as possible from non-condensable gases. Our brochure“Pure steam supply lines for steam sterilization” has been written with the aim of describing the requirements for pure steam supply systems in steam sterilization technology and highlighting possible technical solutions. Steam for sterilization processes should be saturated steam that is free of impurities and contains no non-condensable gases. Impurities in the sterilizing steam enter the autoclave and are transferred to the products. This can damage or contaminate your products, form unwanted deposits or even endanger human life. The residues in the steam can come from the feed water, the steam boiler or the steam lines. Our brochure provides many valuable tips on how to optimize your steam supply system.